Mode: HCTT 4x400KV/1600KVA
Cylinder type Single phase, 50Hz
Rated voltage: Low voltage: 3 KV High voltage: 1600KV
Rated current: LV current: 2*266.7A HV current: 1A
Rated capacity: 1600KVA
Voltage ratio of measuring winding: 3000:1
Waveform distortion: ≤3%
Impedance voltage of three staged in serial connection: ≤25%
Dielectric loss factor: tanδ≤0.3%
PD level: ≤8pC under 100% rated voltage; ≤5pC under 80% rated voltage
Insulation level: self-inducted withstand for 1 min under 110% rated voltage
Voltage unevenness of three stages in serial connection: <±5%
Working time: 1 hour on and 1 hour off under rated voltage and rated current; no more than 8 times per day;
Over current ability: continuous working for 5 min under 120% rated current without any damage and distortion to transformer winding;
Over voltage ability: withstand for 1 min under 110% rated voltage without any damage to transformer insulation; the waveform distortion at this moment is ≤5%;
Among lower winding, transmission winding and upper winding installed 1KV capacitor parallel with protection MOA;
Dimension: 6500mm*2000mm*1700mm
Weight: <61000kg


Lowest stage Test Transformer
HCTT 1600KVA/1200KVA/400KVA/400KV
Cylinder typeSingle phase, 50Hz
Rated capacity: LV winding: 1600KVA HV winding: 400KVA Total winding: 1200KVA
Rated voltage: LV winding: 2*3KV HV winding: 400KV Total winding: 3KV
Rated current: LV winding: 2*266.7A HV winding: 1 A Total winding: 400A
Short-circuit impedance: <5%
Voltage ratio of measuring winding: 1000:1
Waveform distortion: ≤3%
Cooling: ONAN
Dielectric loss factor: tanδ≤0.3%
PD level: ≤5pC under 100% rated voltage; ≤3pC under 80% rated voltage
Winding to ground insulation:
HV winding to LV winding, measuring winding to ground: >2500MΩ
LV winding to HV winding, measuring winding to ground: >2500MΩ
Measuring winding to HV winding, LV winding to ground: >2500MΩ
Insulation level:
LV winding to measuring winding, HV winding to ground withstand: 5KV/MIN
Measuring winding to LV winding, HV winding to ground withstand: 5KV/MIN
HV winding to measuring winding, LV winding to ground withstand: 5KV/MIN
Working time: 1 hour on and 1 hour off under rated voltage and rated current; no more than 8 times per day;
Over current ability: continuous working for 5 min under 120% rated current without any damage and distortion to transformer winding;
Over voltage ability: withstand for 1 min under 120% rated voltage without any damage to transformer insulation; the waveform distortion at this moment is ≤5%;
Between exciter winding and transmission winding installed 3KV capacitor parallel with protection MOA;
Middle/Lower stage Test Transformer
HCTT 1200KVA/800KVA/400KVA/400KV
Cylinder type
Single phase, 50Hz
Rated capacity: LV winding: 800KVA HV winding: 400KVA Total winding: 400KVA
Rated voltage: LV winding: 3KV HV winding: 400KV Total winding: 3KV
Rated current: LV winding: 266.7A HV winding: 1 A Total winding: 133.3A
Short-circuit impedance: <5%
Waveform distortion: ≤3%
Cooling: ONAN
Dielectric loss factor: tanδ≤0.3%
PD level: ≤5pC under 100% rated voltage; ≤3pC under 80% rated voltage
Winding to ground insulation:
HV winding to LV winding, measuring winding to ground: >2500MΩ
LV winding to HV winding, measuring winding to ground: >2500MΩ
Measuring winding to HV winding, LV winding to ground: >2500MΩ
Insulation level:
LV winding to measuring winding, HV winding to ground withstand: 5KV/MIN
Measuring winding to LV winding, HV winding to ground withstand: 5KV/MIN
HV winding to measuring winding, LV winding to ground withstand: 5KV/MIN
Working time: 1 hour on and 1 hour off under rated voltage and rated current; no more than 8 times per day;
Over current ability: continuous working for 5 min under 120% rated current without any damage and distortion to transformer winding;
Over voltage ability: withstand for 1 min under 120% rated voltage without any damage to transformer insulation; the waveform distortion at this moment is ≤5%;
Between exciter winding and transmission winding installed 3KV capacitor parallel with protection MOA;
Middle / Upper stage Test Transformer
HCTT 800KVA/400KVA/400KVA/400KV
Cylinder type
Single phase, 50Hz
Rated capacity: LV winding: 1200KVA HV winding: 400KVA Total winding: 800KVA
Rated voltage: LV winding: 3KV HV winding: 400KV Total winding: 3KV
Rated current: LV winding: 400A HV winding: 1 A Total winding: 266.7A
Short-circuit impedance: <5%
Waveform distortion: ≤3%
Cooling: ONAN
Dielectric loss factor: tanδ≤0.3%
PD level: ≤5pC under 100% rated voltage; ≤3pC under 80% rated voltage
Winding to ground insulation:
HV winding to LV winding, measuring winding to ground: >2500MΩ
LV winding to HV winding, measuring winding to ground: >2500MΩ
Measuring winding to HV winding, LV winding to ground: >2500MΩ
Insulation level:
LV winding to measuring winding, HV winding to ground withstand: 5KV/MIN
Measuring winding to LV winding, HV winding to ground withstand: 5KV/MIN
HV winding to measuring winding, LV winding to ground withstand: 5KV/MIN
Working time: 1 hour on and 1 hour off under rated voltage and rated current; no more than 8 times per day;
Over current ability: continuous working for 5 min under 120% rated current without any damage and distortion to transformer winding;
Over voltage ability: withstand for 1 min under 120% rated voltage without any damage to transformer insulation; the waveform distortion at this moment is ≤5%;
Between exciter winding and transmission winding installed 3KV capacitor parallel with protection MOA;
Uppest stage Test Transformer
HCTT 400KV/400KVA
Cylinder type
Single phase, 50Hz
Rated capacity: LV winding: 400KVA HV winding: 400KVA
Rated voltage: LV winding: 3KV HV winding: 400KV
Rated current: LV winding: 133.3A HV winding: 1 A
Short-circuit impedance: <5%
Waveform distortion: ≤3%
Cooling: ONAN
Dielectric loss factor: tanδ≤0.3%
PD level: ≤5pC under 100% rated voltage; ≤3pC under 80% rated voltage
Winding to ground insulation:
HV winding to LV winding, measuring winding to ground: >2500MΩ
LV winding to HV winding, measuring winding to ground: >2500MΩ
Measuring winding to HV winding, LV winding to ground: >2500MΩ
Insulation level:
LV winding to measuring winding, HV winding to ground withstand: 5KV/MIN
Measuring winding to LV winding, HV winding to ground withstand: 5KV/MIN
HV winding to measuring winding, LV winding to ground withstand: 5KV/MIN
Working time: 1 hour on and 1 hour off under rated voltage and rated current; no more than 8 times per day;
Over current ability: continuous working for 5 min under 120% rated current without any damage and distortion to transformer winding;
Over voltage ability: withstand for 1 min under 120% rated voltage without any damage to transformer insulation; the waveform distortion at this moment is ≤5%;
Between exciter winding and transmission winding installed 3KV capacitor parallel with protection MOA;
HIMALAYAL provides innovative products that comprise high voltage test equipment, full-set high voltage test instruments, on-site test systems and on-line monitoring systems, and services regarding tests, diagnostics and asset monitoring for clients in power industry. We help to ensure safety and reliability in the power generation, transmission and distribution.
Himalayal Corporation Limited
Tel:+86 21 5059 9069
Email: info@himalayal.com
http://www.himalayal.com
http://www.himalayal.com.cn