≤800kV Range
HIVG 100KV-5KJ Impulse Voltage Generator
Home > Systems > Impulse Generator > Impulse Generator > ≤800kV Range

HIVG 100KV-5KJ Impulse Voltage Generator

Lead time: 60 days

Mode: HIVG 100KV-5KJ

Related Standards: 
* IEC 60060-1-2010 High-voltage test techniques - Part 1: General definitions and test requirements
* IEC 60060-2-2010 High-voltage test techniques. Part 2: Measuring systems
* IEC 61083-1-2001 Instruments and software used for measurement in high-voltage impulse tests - Part 1: Requirements for instruments
* IEC 61083-2-1996 Digital recorders for measurements in high-voltage tests - Part 2: Evaluation of software
* ANSI/IEEE 4a-2001 Techniques of High-Voltage Testing
* ANSI/IEEE 376-1975 Measurement of Impulse Strength and Impulse Bandwidth
* ANSI/IEEE 1122-2007 Digit Recorders for Measurements in High-Voltage Impulse Tests
* ASTM D69-D2484 Electrical Insulation ()
* ASTM D2518-Latest Electrical Insulation (
)

impulse voltage generator 100KV impulse voltage generator 100KV

impulse voltage generator 100KV impulse voltage generator 100KV

Technical Parameters: 
1) Rated Voltage: ±100kV
2) Rated stage voltage: ±100kV
3) Rated charging voltage: ±100kV
4) Rated energy: 5kJ
5) Stages: 1 stages
6) Pulse capacitor: 100kV 1uF 1sets totally.
7) Stage energy: 5kJ
8) Impulse voltage waveform parameter:
     a) Under the condition of Load capacitance:400~3400pF (Include Capacitance Voltage Divider)
     b) Standard lightning impulse voltage full-wave
          T1=1.2μS±30%, T2=50μS±20%, oscillation at peak value less than 5% of amplitude
     c) This two impulse voltage waveform parameter and deviation meet the requirement of GB311 and GB16927.
9) Minimum output voltage >20%Un
10) Synchronous range: stage voltage is between 20%~100% of rated voltage.
       Positive and negative polarity synchronous range less than 25%.
11) Synchronous discharge out of control rate <5%
12) Sparking range: 10%-100%Un
13) Charging voltage instability <±1.0%
14) Working time:
       When voltage above than 2/3 rated voltage, the systems can continuous working each 120s (charging and discharging).
       When the voltage is below 2/3 of rated voltage, the system can continuous working each 60s (charging and discharging) 

Features: 
1)easy and fast change of configuration, to satisfy various of testing requirements
2) computer-aided control system, user friendly operation interface
3) excellent safety protection device
4)can be used in combination with AC or DC test system

Principle Diagram: 



  