≤800kV Range
HIVG 300KV-15KJ Impulse Voltage Generator
Home > Systems > Impulse Generator > Impulse Generator > ≤800kV Range

HIVG 300KV-15KJ Impulse Voltage Generator

Lead time: 60 days

Mode: HIVG 300KV-15KJ

Related Standards: 
* IEC 60060-1-2010 High-voltage test techniques - Part 1: General definitions and test requirements
* IEC 60060-2-2010 High-voltage test techniques. Part 2: Measuring systems
* IEC 61083-1-2001 Instruments and software used for measurement in high-voltage impulse tests - Part 1: Requirements for instruments
* IEC 61083-2-1996 Digital recorders for measurements in high-voltage tests - Part 2: Evaluation of software
* ANSI/IEEE 4a-2001 Techniques of High-Voltage Testing
* ANSI/IEEE 376-1975 Measurement of Impulse Strength and Impulse Bandwidth
* ANSI/IEEE 1122-2007 Digit Recorders for Measurements in High-Voltage Impulse Tests
* ASTM D69-D2484 Electrical Insulation ()
* ASTM D2518-Latest Electrical Insulation (
)

Technical Parameters: 
1) Rated Voltage: ±300kV
2) Rated stage voltage: ±100kV
3) Rated charging voltage: ±100kV
4) Rated energy: 15kJ
5) Stages: 3 stages
6) Pulse capacitor: 100kV 1uF 3sets totally.
7) Stage energy: 5kJ
8) Impulse voltage waveform parameter:
    a) Under the condition of Load capacitance:300~3000pF
    b) Standard lightning impulse voltage full-wave:
        T1=1.2μS±30%, T2=50μS±20%, oscillation at peak value less than 5% of amplitude
    c) Standard lightning impulse chopped wave: Chopped time: 2~5μS; Peak deviation: ≤ 3%;
    d) This two impulse voltage waveform parameter and deviation meet the requirement of GB311 and GB16927.
9) Minimum output voltage >20%Un
10) Voltage utilization factor: Standard lightning impulse wave utilization factor≥90%, when load capacitance less than 1000pF; Standard lightning impulse wave utilization factor ≥85%, when load capacitance less than 1000pF;
11) Synchronous range: stage voltage is between 20%~100% of rated voltage.
       Positive and negative polarity synchronous range less than 25%.
12) Synchronous discharge out of control rate <5%
13) Sparking range: 10%-100%Un
14) Charging voltage instability <±1.0%
15) Working time:
       When voltage above than 2/3 rated voltage, the system can continuous working each120s (charging and discharging).
       When the voltage is below 2/3 of rated voltage, the system can be continuous working each 60s (charging and discharging).