800kV ~ 1600kV Range
HIVG 1000KV-100KJ Impulse Voltage Generator
Home > Systems > Impulse Generator > Impulse Generator > 800kV ~ 1600kV Range

HIVG 1000KV-100KJ Impulse Voltage Generator

Lead time: 75days

Mode: HIVG 1000KV-100KJ

Related Standards: 
* IEC 60060-1-2010 High-voltage test techniques - Part 1: General definitions and test requirements
* IEC 60060-2-2010 High-voltage test techniques. Part 2: Measuring systems
* IEC 61083-1-2001 Instruments and software used for measurement in high-voltage impulse tests - Part 1: Requirements for instruments
* IEC 61083-2-1996 Digital recorders for measurements in high-voltage tests - Part 2: Evaluation of software
* ANSI/IEEE 4a-2001 Techniques of High-Voltage Testing
* ANSI/IEEE 376-1975 Measurement of Impulse Strength and Impulse Bandwidth
* ANSI/IEEE 1122-2007 Digit Recorders for Measurements in High-Voltage Impulse Tests
* ASTM D69-D2484 Electrical Insulation ()
* ASTM D2518-Latest Electrical Insulation (
)
 

impulse voltage generator 1000KV impulse voltage generator 1000KV

impulse voltage generator 1000KV impulse voltage generator 1000KV

Technical Parameters: 
1) Rated Voltage: ±1000kV
2) Rated stage voltage:100KV
3) Rated energy: 100kJ
4) Rated impulse capacity: 0.2uf                    
5)Stages:10 stages
6) Rated stage capacity/ stage voltage/ stage energy: 2uF/100KV/10KJ
7) Synchronous range: >20%
8) Synchronous discharge out of control rate <2%
9) Minimum output voltage: >10%UH
10) Charging voltage instability <±1.0%
11) Sparking range: 10%-100%UH
12) Working time:
       When voltage above than 70% rated voltage, the systems can continuous working each 120s (charging and discharging).
       When the voltage is below 70% of rated voltage, the system the system can continuous working each 60s(charging and discharging).
13) Impulse voltage waveform parameter:
       a) Standard lightning impulse voltage full-wave: 1.2±30%ms /50±20%ms; Overshoot: <5%   Efficiency: >90%
       b) Standard switching impulse wave: 250±20%ms /2500±60%ms, efficiency is >80%
       c) These two impulse voltage waveform parameter and deviation meet the requirement of standards GB311 and GB16927 about the impulse voltage and impulse current waveform parameters.


Principle Diagram: 

  