Summary of facility 

This equipment will be used for carrying out the residual voltage test and impulse current withstand test of the arrester element.
Impulse current generator Marx generator

2.Specification 

nApplied standard:IEC60099-4(2014) Metal-oxide surge arresters without gaps for a.c systems

nLightning impulse residual voltage test, 

nHigh current impulses

3.Specification

nCurrent wave form: 8/20us, 4/10us

nCurrent peak value (max): 20kA(8/20us), 120kA(4/10us)

nResidual voltage (max): 20kV / 20kA @8/20us, 28kV / 100kA@4/10us

nCurrent transformer: 200kA / 500V (250V with terminator) @4/10us 100kA 

    50kA / 500V (250V with terminator) @ 8/20us 20kA

nVoltage divider: Resistive divider 300:1 for residual voltage test,high current withstand test

nMeasuring device: Digital oscilloscope, 500MHz, 4ch (Tektronix)

nPower source: 3 phase 380V/400V (for unit Loads), and 1 phase 230V;  Power frequency is 50Hz/60Hz

nExternal interlock: Dry contact ( test ready is on )

nCircuit change: Connection change of L and C can be easily.

nCompartment: The high voltage power supply portion is at all times isolated test sample portion is surrounded by insulating material.

nControl: Imp circuit charging voltage , Discharge voltage
Impulse current generator Marx generator

4.Measuring item: [Residual voltage test,High current impulse test]

nResidual voltage

nDischarge current

nEnergy(calculation)

nCurrent waveform(Tf,Tt) 
Impulse current generator Marx generator

5.Production type: Metal Oxide Varistor (MOV)

The test cabinet will be customized according to object dimension.
Impulse current generator Marx generator

  